The quality requirements for nitrogen hydrogen air integrated machines mainly include the following aspects:
Gas purity: The purity of nitrogen should reach 99.999% or higher, and the purity of hydrogen should also reach 99.999%. The air should be dry, oil-free, and undergo three levels of purification (silica gel, molecular sieve, activated carbon) to achieve a cleanliness level of iso 8573-1 class 1 standard.
Flow and pressure: The output flow range of nitrogen and hydrogen is 0-300ml/min to 0-500ml/min, and the output flow range of air is 0-2000ml/min to 0-5000ml/min. In terms of output pressure, the output pressure of nitrogen, hydrogen, and air is usually 0.4-0.5mpa, which supports customized adjustment.
Intelligent control: Equipped with a fully automatic constant voltage and current system, automatic flow tracking and adjustment, and equipped with LED digital display for real-time parameters. Built in microcontroller control, with multiple protection functions such as overvoltage, leakage, and abnormal temperature.

Safety performance: Using stainless steel gas storage tanks with anti backflow technology to ensure zero leakage of electrolyte and avoid contaminating the detector. The air section is equipped with two-stage overvoltage protection devices, two-stage voltage stabilization devices, automatic drainage devices, and dust filtration devices to ensure safety and reliability.
Environmental adaptability: Suitable for environments with a temperature range of 10-40 ℃ and a relative humidity of ≤ 85%, without a large amount of dust and corrosive gases.
Convenience of maintenance: Equipped with intelligent drainage devices that automatically drain condensed water, reducing manual maintenance. Stable gas production without attenuation, only distilled water needs to be added for regular use.